Cor Vasa 2020, 62(1):80-84 | DOI: 10.33678/cor.2019.011

Pacing-induced right ventricular cardiomyopathy resynchronized using His bundle pacing

Martin Kameník, Pavel Osmančík, Petr Štros, Dalibor Heřman, Jana Veselá, Radka Procházková, Karol Čurila
Cardiocenter, Third Faculty of Medicine, Charles University and Faculty Hospital Kralovske Vinohrady, Prague

Permanent unifocal right ventricular pacing can lead to ventricular dyssynchrony with subsequent dysfunction of the left ventricle, which can lead to heart failure. Heart failure caused by right ventricular pacing was traditionally associated with pacing from the apex of the right ventricle. Over the last two decades, significant efforts have been made to find alternative pacing sites that can reduce or prevent the risk of heart failure. Based on the latest findings, pacing the His bundle area appears extremely promising. In this case report, we present a 79-year-old male, who developed left ventricular dysfunction with heart failure within six months of right ventricular septal pacing. The patient was upgraded to CRT-D, and one of the pacing leads was implanted in the His bundle area. Selective His bundle pacing led to normalization of QRS duration and a significant improvement of heart failure symptoms. Echocardiography performed after three months showed normalized left ventricular function.

Keywords: Heart failure, His bundle pacing, Pacemaker induced cardiomyopathy, Physiological pacing

Received: March 10, 2019; Accepted: March 24, 2019; Published: February 28, 2020  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Kameník M, Osmančík P, Štros P, Heřman D, Veselá J, Procházková R, Čurila K. Pacing-induced right ventricular cardiomyopathy resynchronized using His bundle pacing. Cor Vasa. 2020;62(1):80-84. doi: 10.33678/cor.2019.011.
Download citation

References

  1. Sweeney M, Hellkamp A, Ellenbogen K, et al. Adverse Effect of Ventricular Pacing on Heart Failure and Atrial Fibrillation Among Patients With Normal Baseline QRS Duration in a Clinical Trial of Pacemaker Therapy for Sinus Node Dysfunction. Circulation 2003;107:2932-2937. Go to original source... Go to PubMed...
  2. Tse H, Lau C. Long-Term Effect of Right Ventricular Pacing on Myocardial Perfusion and Function. J Am Coll Cardiol 1997;29:744-749. Go to original source... Go to PubMed...
  3. Vijayaraman P, Bordachar P, Ellenbogen K. The Continued Search for Physiological Pacing. J Am Coll Cardiol 2017;69:3099-3114. Go to original source... Go to PubMed...
  4. Deshmukh P, Casavant D, Romanyshyn M, Anderson K. Permanent direct His bundle pacing: a novel approach to cardiac pacing in patients with normal His-Purkinje activation. Circulation 2000;101:869-877. Go to original source... Go to PubMed...
  5. Zanon F, Svetlich C, Occhetta E, et al. Safety and Performance of a System Specifically Designed for Selective Site Pacing. Pacing Clin Electrophysiol 2010;34:339-347. Go to original source... Go to PubMed...
  6. Vijayaraman P, Naperkowski A, Subzposh F, et al. Permanent His-bundle pacing: Long-term lead performance and clinical outcomes. Heart Rhythm 2018;15:696-702. Go to original source... Go to PubMed...
  7. Abdelrahman M, Subzposh F, Beer D, et al. Clinical Outcomes of His Bundle Pacing Compared to Right Ventricular Pacing. J Am Coll Cardiol 2018;71:2319-2330. Go to original source... Go to PubMed...
  8. Sharma P, Dandamudi G, Herweg B, et al. Permanent His bundle pacing as an alternative to biventricular pacing for cardiac resynchronization therapy: A multicenter experience. Heart Rhythm 2018;15:413-420. Go to original source... Go to PubMed...

This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (CC BY-NC 4.0), which permits non-comercial use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.