Cor Vasa 2022, 64(4):411-422 | DOI: 10.33678/cor.2022.083
Feasibility of evaluation of Polar H10 chest-belt ECG in patients with a broad range of heart conditions
- Department of Internal Medicine I - Cardiology, Palacky University and University Hospital Olomouc, Olomouc
Background: The chest-belt can be used to obtain a 1-lead ECG. Data from it have been validated for the determination of heart rate and for the possibility to detect atrial fibrillation (AF) compared to ECG-Holter on a short ECG recording in selected patients. However, validation of the possibility to evaluate long ECG recordings in patients with a wide range of heart diseases has not yet been performed. Methodology and results: 54 hospitalized patients, 53 outpatients and 54 healthy controls were enrolled in the study (n = 161 in total). Using a Polar H10 chest-belt, 1-2 hours of ECG were recorded in all patients (1 153 229 heartbeats, average heart rate 76.6/min, 86.3% in sinus rhythm, 13.7% with atrial fibrillation, 0.46% atrial premature beats, 0.49% ventricular premature beats). The presence of noise was 2.16% (A: 2.31%; B: 1.95%; C: 2.20%). 1 128 319 /1 153 229 were evaluated as easy to interpret. Using ECG from the belt, the basic rhythm was reliably determined by the physician in majority of patients (51/54, 94.4% in hospitalized patients; in 100% of outpatients and healthy controls) when compared to 12-lead ECG. 3 cases were evaluated as unclear; in all of these cases, all QRS complexes were stimulated by a pacemaker. In hospitalized patients, real-time ECG from the belt was comparable to telemetric ECG monitoring (match in 53/54, 98.1%).
Conclusion: The ECG obtained from the chest-belt in hospitalized patients and outpatients with a wide range of cardiovascular diseases, as well as in healthy individuals, is usable in real practice for evaluation of baseline rhythm, atrial fibrillation and premature contractions with a minimal proportion of difficulties to interpret recordings due to artefacts. Caution should be exercised in interpretation of the ECG in patients with stimulated rhythm and in patients with atrial flutter. The chest belt can be used as a means for continuous monitoring of ECG, evaluation of rhythm and screening of atrial fibrillation.
Keywords: Atrial fibrillation, Chest-belt, Noise, Polar H10, Screening tools
Received: July 15, 2022; Revised: July 15, 2022; Accepted: July 15, 2022; Prepublished online: August 10, 2022; Published: September 1, 2022 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- Kaptoge S, Pennells L, De Bacquer D, et al. World Health Organization cardiovascular disease risk charts: revised models to estimate risk in 21 global regions. Lancet Glob Health 2019;7:e1332-e1345.
Go to original source...
Go to PubMed...
- Hindricks G, Potpara, T, Dagres N, et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS) The Task Force for the diagnosis and management of atrial fibrillation of the European Society of Cardiology (ESC) Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESC. Eur Heart J 2021;42:373-498.
Go to original source...
Go to PubMed...
- Benjamin EJ, Muntner P, Alonso A, et al.; American Heart Association Council on Epidemiology and Prevention Statistics Committee and Stroke Statistics Subcommittee. Heart disease and stroke statistics 2019 update: a report from the American Heart Association. Circulation 2019;139:e56-e528.
Go to original source...
Go to PubMed...
- Chugh SS, Havmoeller R, Narayanan K, et al. Worldwide epidemiology of atrial fibrillation: a Global Burden of Disease 2010 Study. Circulation 2014;129:837-847.
Go to original source...
Go to PubMed...
- Wynn GJ, Todd DM, Webber M, et al. The European Heart Rhythm Association symptom classification for atrial fibrillation: validation and improvement through a simple modification. Europace 2014;16:965-972.
Go to original source...
Go to PubMed...
- Steinberg JS, O'Connell H, Li S, Ziegler PD. Thirty-second gold standard definition of atrial fibrillation and its relationship with subsequent arrhythmia patterns: analysis of a large prospective device database. Circ Arrhythm Electrophysiol 2018;11:e006274.
Go to original source...
Go to PubMed...
- Mairesse GH, Moran P, Van Gelder IC, et al.; ESC Scientific Document Group. Screening for atrial fibrillation: a European Heart Rhythm Association (EHRA) consensus document endorsed by the Heart Rhythm Society (HRS), Asia Pacific Heart Rhythm Society (APHRS), and Sociedad Latinoamericana de Estimulacion Cardiaca y Electrofisiologia (SOLAECE). Europace 2017;19:1589-1623.
Go to original source...
Go to PubMed...
- Gorenek B, Boriani G, Dan GA, et al.; ESC Scientific Document Group. European Heart Rhythm Association (EHRA) position paper on arrhythmia management and device therapies in endocrine disorders, endorsed by Asia Pacific Heart Rhythm Society (APHRS) and Latin American Heart Rhythm Society (LAHRS). Europace 2018;20:895-896.
Go to original source...
Go to PubMed...
- Camm AJ, Simantirakis E, Goette A, et al.. Atrial high-rate episodes and stroke prevention. Europace 2017;19:169-179.
Go to original source...
Go to PubMed...
- Sardana HK, Kanwade R, Tewary S. Arrhythmia detection and classification using ECG and PPG techniques: A review. Phys Eng Sci Med 2021;44:1027-1048.
Go to original source...
Go to PubMed...
- Turakhia MP, Desai M, Hedlin H, et al. Rationale and design of a large-scale, app-based study to identify cardiac arrhythmias using a smartwatch: the Apple Heart Study. Am Heart J 2019;207:66-75.
Go to original source...
Go to PubMed...
- Halcox JPJ, Wareham K, Cardew A, et al. Assessment of remote heart rhythm sampling using the AliveCor heart monitor to screen for atrial fibrillation: the REHEARSE-AF Study. Circulation 2017;136:1784-1794.
Go to original source...
Go to PubMed...
- Gilgen-Ammann R, Schweizer T, Wyss T. 2019. RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise. Eur J Appl Physiol 2019;119:1525-1532.
Go to original source...
Go to PubMed...
- Lown M, Yue AM, Shah BN, et al. Screening for atrial fibrillation using economical and accurate technology (from the safety study). Am J Cardiol 2018;122:1339-1344.
Go to original source...
Go to PubMed...
- Hartikainen S, Lipponen JA, Hiltunen P, et al. Effectiveness of the chest strap electrocardiogram to detect atrial fibrillation. Am J Cardiol 2019;123:1643-1648.
Go to original source...
Go to PubMed...
- Barbosa MP, da Silva NT, de Azevedo FM, et al. Comparison of Polar(R) RS800G3 heart rate monitor with Polar(R) S810i and electrocardiogram to obtain the series of RR intervals and analysis of heart rate variability at rest. Clin Physiol Funct Imaging 2016;36:112-117.
Go to original source...
Go to PubMed...
- Gamelin FX, Berthoin S, Bosquet L (2006) Validity of the polar S810 heart rate monitor to measure R-R intervals at rest. Med Sci Sports Exerc 38(5):887-893.
Go to original source...
Go to PubMed...
- Giles D, Draper N, Neil. Validity of the Polar V800 heart rate monitor to measure RR intervals at rest. Eur J Appl Physiol 2016;116:563-571.
Go to original source...
Go to PubMed...
- Costa RBMPd, Mícolis AF, Marcelo PC, Marques VLC. Comparison of Polar® RS800G3™ heart rate monitor with Polar® S810i™ and electrocardiogram to obtain the series of RR intervals and analysis of heart rate variability at rest. Clin Physiol Funct Imaging 2016;36:112-117.
Go to original source...
Go to PubMed...
- Caminal P, Sola F, Gomis P, et al. Validity of the Polar V800 monitor for measuring heart rate variability in mountain running route conditions. Eur J Appl Physiol 2018;118:669-677.
Go to original source...
Go to PubMed...
- Cassirame J, Vanhaesebrouck R, Chevrolat S, Mourot L. Accuracy of the Garmin 920 XT HRM to perform HRV analysis. Australas Phys Eng Sci Med 2017;40(4):831-839.
Go to original source...
Go to PubMed...
- Kingsley M, Lewis MJ, Marson RE. Comparison of Polar 810 s and an ambulatory ECG system for RR interval measurement during progressive exercise. Int J Sports Med 2005;26:39-44.
Go to original source...
Go to PubMed...
- Cheatham SW, Kolber MJ, Ernst MP. Concurrent validity of resting pulse-rate measurements: a comparison of 2 smartphone applications, the polar H7 belt monitor, and a pulse oximeter with bluetooth. J Sport Rehabil 2015;24:171-178
Go to original source...
Go to PubMed...
- Plews DJ, Scott B, Altini M, et al. Comparison of heart-rate-variability recording with smartphone photoplethysmography, polar H7 chest strap, and electrocardiography. Int J Sports Physiol Perform 2017;12:1324-1328.
Go to original source...
Go to PubMed...
- Ruha A, Sallinen S, Nissilä S. A real-time microprocessor QRS detector system with a 1-ms timing accuracy for the measurement of ambulatory HRV. IEEE Trans Bio-Med Eng 1997;44:159-167.
Go to original source...
Go to PubMed...
- Parrado E, Garcia MA, Ramos J, et al.Comparison of omega wave system and polar S810i to detect R-R intervals at rest. Int J Sports Med 2010;31:336-341.
Go to original source...
Go to PubMed...
- Task Force of the European Society of Cardiology, the North American Society of Pacing and Electrophysiology. Heart Rate Variability: standards of measurement, physiological interpretation and clinical use. Circulation 1996;93:1043-1065.
- Kearley K, Selwood M, Van den Bruel A, et al. Triage tests for identifying atrial fibrillation in primary care: a diagnostic accuracy study comparing single-lead ECG and modified BP monitors. BMJ Open 2014;4:e004565.
Go to original source...
Go to PubMed...
- Doliwa PS, Frykman V, Rosenqvist M. Short-term ECG for out of hospital detection of silent atrial fibrillation episodes. Scand Cardiovasc J 2009;43:163-168.
Go to original source...
Go to PubMed...
- Sarkar S, Ritscher D, Mehra R. A detector for a chronic implantable atrial tachyarrhythmia monitor. IEEE Trans Biomed Eng 2008;55:1219-1224.
Go to original source...
Go to PubMed...
- Lake DE, Moorman JR. Accurate estimation of entropy in very short physiological time series: the problem of atrial fibrillation detection in implanted ventricular devices. Am J Physiol Heart Circ Physiol 2011;300:H319-H325.
Go to original source...
Go to PubMed...
This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (CC BY-NC 4.0), which permits non-comercial use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.